| Functional principle | Ultrasonic distance measurement |
| Operating range | 350–3,400 mm |
| Limiting range | 5,000 mm; distinct from operating range |
| Detection area | Depends on reflection, size and target alignment; reference diagrams use aligned 500 × 500 mm plate and Ø27 mm pipe |
| Resolution | ≥0.18 mm |
| Repeatability | ±0.15% of current measured value; minimum value ≥resolution |
| Measurement accuracy | ±1% of current measured value |
| Temperature compensation | Available; can be disabled; optimum after approximately 30 min |
| Uncompensated temperature drift | 0.17%/K with temperature compensation disabled |
| Response time | 180 ms; subsequent analog smoothing may increase response time by up to 200% |
| Output time | 43 ms |
| Ultrasonic frequency | Typical 120 kHz; acoustic frequency |
| Analog output | 1 × current/voltage output; automatic 4–20 mA or 0–10 V selection according to connected load |
| Analog load requirements | 4–20 mA: ≤500 Ω, reduced to ≤100 Ω at supply ≤20 V; 0–10 V: ≥100,000 Ω |
| Analog output resolution | 12 bit |
| Adjustment | Local teach-in and numeric analog scaling; rising/falling output characteristic; measurement filters, foreground suppression, sensitivity/detection area; display deactivation and factory reset |
| Synchronization and multiplexing | Via MF pin 5; up to 50 interconnected sensors; multiplexing requires distinct addresses |
| Multifunction input | 1 × MF |
| Multiple-sensor separation | Without synchronization/multiplexing: parallel >200 cm; opposite >1,800 cm |
| Power supply | 9–30 V DC; 15–30 V DC required for analog voltage output; limit values; reverse-polarity protected |
| Supply network requirement | Short-circuit protected network, maximum 8 A, Class 2 |
| Power consumption | ≤2.4 W without load |
| Initialization time | <300 ms |
| Indicators and controls | LED display, 2 status LEDs, local T1/T2 controls; green analog LED indicates object within scaling range |
| Connection | M12 five-pin male connector |
| Pin assignment | 1 BN=L+; 3 BU=M/0 V; 2 WH=QA analog output; 5 GY=Sync/Multi and Connect+ communication; 4 BK=Q/inverted Q digital output |
| Mounting thread | M30 × 1.5 |
| Mounting nuts | 36 mm across flats; maximum tightening torque 40 Nm |
| Sending axis | Straight |
| Housing materials | Nickel-plated brass and PBT; display TPU |
| Transducer materials | Polyurethane foam and glass epoxy resin |
| IP rating | IP65/IP67 |
| Protection class | III |
| Operating temperature | −25 to +70 °C |
| Storage temperature | −40 to +85 °C |
| MTTFD | 101 years |
| DCavg | 0% |
| Physical Weight (kg) | 0.21 |
| Digital output | 1 × PNP; maximum output current ≤200 mA |
| Switching frequency | 4 Hz |
| Hysteresis | 50 mm; configurable using Connect+ |
| Switching modes | Switching point (DtO), window, object between sensor and background (ObSB); output inversion and on delay |